Authors: | T. Claes, W. Bogaerts, P. Bienstman | Title: | Experimental characterization of a silicon photonic biosensor consisting of two cascaded ring resonators based on the Vernier-effect and introduction of a curve fitting method for an improved detection limit | Format: | International Journal | Publication date: | 10/2010 | Journal/Conference/Book: | Optics Express
| Volume(Issue): | 18(22) p.22747--22761 | DOI: | 10.1364/oe.18.022747 | Citations: | 263 (Dimensions.ai - last update: 24/11/2024) 130 (OpenCitations - last update: 3/5/2024) Look up on Google Scholar
| Download: |
(2MB) |
Abstract
We studied a refractive index sensor that consists of two cascaded ring resonators and that works analogously to a Vernier-scale. We implemented it in silicon-on-insulator and experimentally determined its sensitivity to be as high as 2169nm/RIU in aqueous environment. We derived formulas describing the sensor’s operation, and introduced a fitting procedure that allows to accurately detect changes in the sensor response. We determined the detection limit of this first prototype to be 8.3*10−6RIU. Related Research Topics |
|
|
Citations (OpenCitations)
|
|